Image Volume Fractionator workflow >
5. Define SRS grid layout
Overview
In this step, the SRS (systematic random sampling) grid size defines how close your counting frames are to each other.
- Smaller grid: You visit more sites within the traced contour, and count more cells or other objects.
- Larger grid: You visit fewer sites (this may increase your coefficient of error and obtain a less-precise estimate).
Procedure
Define SRS grid layout
-
Choose one of the following ways to define the SRS grid layout:
Grid size and counting frame size remain consistent throughout all of the sections for a given region of interest. The number of sites, however, may change because the grid is applied randomly every time you start a new probe run. Click Display Changes or Estimate Grid Size repeatedly to see how the grid is applied differently every time.
-
Manually enter the grid size
- Type in values (µm) for X and Y grid spacing.
- Click Display Changes.
-
Enter a percentage of the region of interest that you want to sample.
-
Type the percent to sampled in the Percentage box.
-
Click Display Changes
-
-
Determine grid size based on number of sampling sites:
-
Enter the desired number of sites in the Approximate Sites box.
-
Click Estimate Grid Size.
This option is ideal if you don’t know your desired X and Y grid spacing. For instance, if you want to count a mean of 10 sites based on the current representative section, set the grid size to include approximately 10 counting frames in the region of interest.
-
-
-
We recommend that you check the box to Round to increments of 1 micron or larger.
Undo my changes: Click to remove your input for this step and revert to the default values.