Troubleshooting
Lens calibration failed: The quick line measurement was significantly off.
OR
Parcentric/parfocal calibration failed: When parcentric/parfocal was tested, the tracing did not remain in the same position between objectives.
Try the following:
- Check the camera alignment.
- Check the stage movements.
Once those have all been examined, perform grid-tuning and parcentric/parfocal calibration again to see if any issues were resolved:
- Re-calibrate (grid tune) the lenses and test the calibration with the quick measure line (see Calibrate lenses with calibration grids ).
- Re-calibrate parcentric and parfocal and test the calibration by drawing a contour and switching through the objectives to make sure it stays aligned (see Parcentric/Parfocal calibration ).
If, after checking the camera, stage, and Z encoder, the calibration remains unsatisfactory, contact us for additional assistance.