Camera histogram, Background correction for brightfield modality
This is an ideal histogram for most types of brightfield data. The dynamic range is nearly the entire range of the camera. |
|
|
The dynamic range is too low.
To improve the dynamic range, you may move the white point slider to the left (see figure on the left). This may give the illusion of a better image. However, most of the dynamic range is lost (see 2nd histogram) and the signal-to-noise ratio is lower. |
|
Background image only
Good example for the acquisition of a brightfield background image -- Capturing a background image for slide scanning(brightfield only)
This histogram indicates the presence of dust or other artifacts in the background image.
|
The red line on the right of the graph indicates saturation. If the image is completely saturated, the function (represented in gray) is not visible; only the red line appears. If enabled, Clip Detect will show most of the image as red.
The blue line on the left indicates that the signal is below the range. |
|
This histogram suggests that the light source is not sufficient the camera. To correct this, examine an image in the eyepieces.
The blue line on the left may not appear depending on the sensitivity of the camera. |