You can also use Move>AutoMove to enable AutoMove.
Repositions the cursor to the center when AutoMove is used.
Use the center of the AutoMove area instead of the default (center of the tracing window) when repositioning the stage.
Use to control how the stage is moved (or to navigate acquired images) when using the Move>field moves arrows or Move>Meander Scan.
The amount of movement can be set to a fixed percentage of the size of the tracing window.
Set the Field Movement Size visually.
Unlike the previous options, the values entered might not be applicable to all lenses because they are absolute values.
Change the color of the box.
The default action for the mouse wheel is to adjust the cursor size. Also see Using the mouse wheel.
Check to change the mouse wheel action to focusing.
Each moves through the Z-axis by the distance specified in Z distance per wheel click.
Use for frequent adjustments of the Z distance per wheel click.
To navigate one plane at a time, select 0.5 from the State drop-down menu.
To enable both cursor size adjustment and focusing with the mouse wheel:
Available if Parcentric/Parfocal calibration has been performed.
Corrections are automatically applied whenever you change lenses.
The slide scanning module is required for this option to be visible.
You can define the Z range for each lens to be used with the slide scanning module. The higher power the objective, the smaller the Z range and the finer the step size.
Then Autofocus is applied 20 (10*2) microns above and below Z.
If Autofocus is enabled in the lens preferences, you can also set the autofocus at the time of a slide scan: Click the Set Autofocus Range button in the Slide scanning setup window (see Settings>Focus Adjustments).