Define image stack settings
SRS image stack workflow — Step 8
Collect the following set of images
Collect additional images above | below the selection: Designate the desired distance in microns.
Collect Images for:
- Mounted thickness: Acquires the image stack for the entire specified mounted thickness.
- Probe and top guard zone: Acquires the image stack for the specified disector height (probe) and top guard zone only; it will not acquire the bottom guard zone.
- Probe only: Acquires the image stack for the specified disector height (probe) ONLY; it will not acquire the top and bottom guard zones.
Enter Image Stack height information
Top guard zone / Probe height: Must be less than the mounted thickness when combined.
Mounted thickness: Must be less than the section cut thickness.
If you don't know the exact mounted thickness and wish to determine it manually, click Set. The Measure tissue thickness is displayed.
- Focus to the bottom of the tissue and click Set bottom of section.
-
Focus to the top of the tissue and click Set top of section.
- The height of the tissue is displayed; accept the measurement by clicking Add measurement to list. This sets the height as your Mounted thickness.
- If the measurement is unsatisfactory, click Delete measurement and start again, or click Revisit measurement to edit your current settings.
Distance Between Images
This is the step size of your stack.